Signal Generator Settings
R&S
®
SGT100A
162
User Manual 1176.8674.02 ─ 07
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:ENVelope:OFFSet
DC Modulator characteristics
Refer to the product documentation of the external DC modulator for information on its
characteristics.
The following settings are required:
EMF ← DC Modulator characteristics
Defines whether the EMF or the voltage value is displayed.
An EMF-based calculation assumes an open-end circuit. Disable this parameter for
testing in more realistic conditions, where you define the input impedance of the used
external DC modulator
and the instrument calculates the envelope output voltage
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:ENVelope:EMF[:STATe]
R
in
← DC Modulator characteristics
With disabled parameter
in
of the used external DC
modulator.
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:ENVelope:RIN
Termination ← DC Modulator characteristics
If the "
> Differential", defines the way the inputs of the DC modulator
are terminated.
"To Ground"
"Wire to Wire"
*
)
*
)
Both inputs of the DC modulator are terminated to
ground.
This termination is also referred as a common mode
voltage.
Generation of Envelope Tracking Signals