Remote control commands
R&S
®
RTO6
1454
User Manual 1801.6687.02 ─ 05
Table 24-4: Header data
Position
Meaning
Example
1
XStart in s
-9.477E-008 = - 94,77 ns
2
XStop in s
9.477E-008 = 94,77 ns
3
Record length of the waveform in Samples
200000
4
Number of values per sample interval. For reference
waveforms the number is always 1.
1
Suffix:
<m>
.
1..4
Reference waveform
Example:
REFC:DATA:HEAD?
-9.477E-008,9.477E-008,200000,1
Usage:
Query only
REFCurve<m>:DATA[:VALues]?
Returns the data of the channel waveform points for transmission from the instrument
to the controlling computer. The data can be used in MATLAB, for example.
.
You can retrieve only Y-values, or X- and Y-values. Use
to define this.
Suffix:
<m>
.
1..4
Reference waveform
Return values:
<Data>
List of values according to the format and content settings.
Usage:
Query only
24.10.2.4
Import of multichannel waveform data
Commands for saving waveform data to file are described in
Commands for reference waveforms are listed below.
............................................................................. 1455
REFCurve<m>:MULTichannel:IMPort
.............................................................................1455
............................................................................. 1455
Waveform analysis