Remote control commands
R&S
®
RTO6
1655
User Manual 1801.6687.02 ─ 05
EXPort:WAVeform:INCXvalues
<IncHorValues>
Includes horizontal values in the retrieved data (time or frequency values, depending
on the waveform). X and Y-values are written alternately to the file. If disabled, only Y-
values - mostly voltage values - are written. The X-values are always returned in 64 bit
real format, regardless of the defined data format.
The setting is not available for the export of raw data.
The command affects the content of export files as well as data retrieved with:
●
CHANnel<m>[:WAVeform<n>]:DATA[:VALues]?
●
CALCulate:MATH<m>:DATA[:VALues]?
●
●
●
BUS<m>:PARallel:DATA[:VALues]?
●
●
Parameters:
<IncHorValues>
ON | OFF
*RST:
OFF
Example:
See:
"Exporting interleaved x/y data of a single waveform to CSV file"
"Exporting interleaved x/y data of a zoom to CSV file"
Usage:
Asynchronous command
EXPort:WAVeform:RAW
<RawValues>
Enables the export of raw sample data, and sets the data format to integer 8 bit. In
high definition acquisition mode, the data format is integer 16 bit. Fot INT16, you can
set the byte order using the
command.
The raw format reduces the file size but changes also the precision of the values.
Currently, the setting is not available for the export of digital channel data and data of
R&S
RT-ZVC channels.
The raw export of interleaved X/Y values is not supported.
Parameters:
<RawValues>
ON | OFF
*RST:
OFF
Example:
See:
"Exporting raw data of a single waveform to BIN file"
"Exporting raw data of a measurement gate to BIN file"
Data management