![R&S RTO6 Скачать руководство пользователя страница 1292](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461292.webp)
Remote control commands
R&S
®
RTO6
1292
User Manual 1801.6687.02 ─ 05
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
).
Usage:
Event
24.6
General remote settings
This chapter describes commands that have effect on many other remote commands
in different applications of the instrument.
.............................................................................................1294
............................................................................... 1295
...............................................................................1295
FORMat[:DATA]
<Format>, [<Length>]
Selects the data type that is used for transmission of data from analog channels, math
and reference waveforms, and some measurement results from the instrument to the
controlling computer.
The command sets the data format for the following query commands:
●
CHANnel<m>[:WAVeform<n>]:DATA[:VALues]?
●
CALCulate:MATH<m>:DATA[:VALues]?
●
●
●
●
MEASurement<m>:TRACk:DATA[:VALues]?
Parameters:
<Format>,[<Length>] ASCii | REAL,32 | INT,8 | INT,16
ASCii
Data values are returned in ASCII format as a list of comma-
separated values in floating point format. The length can be
omitted. It is 0 which means that the instrument selects the num-
ber of digits to be returned. The query returns both values (
ASC,
0
).
General remote settings