Remote control commands
R&S
®
RTO6
1410
User Manual 1801.6687.02 ─ 05
SHORter
Triggers on pulses shorter than the given width.
LONGer
Triggers on pulses longer than the given width.
*RST:
WITHin
Usage:
Asynchronous command
TRIGger<m>:PATTern:WIDTh[:WIDTh]
<Width>
For the ranges "Within" and "Outside" (defined using
), the width defines the center of a range which is defined by the limits "
±
Delta"
TRIGger<m>:PATTern:WIDTh:DELTa
For the ranges "Shorter" and "Longer", the width defines the maximum and minimum
pulse width, respectively.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Width>
Range:
100E-12 to 10000
Increment: 100E-9
*RST:
5E-9
Default unit: s
Usage:
Asynchronous command
TRIGger<m>:PATTern:WIDTh:DELTa
<WidthDelta>
Defines a range around the width value specified using
.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<WidthDelta>
Range:
0 to 432
Increment: 500E-12
*RST:
0
Default unit: s
Usage:
Asynchronous command
24.9.13
Serial pattern trigger
The serial pattern trigger is only available for the A-event (Suffix = 1).
Trigger