Remote control commands
R&S
®
RTO6
1409
User Manual 1801.6687.02 ─ 05
TRIGger<m>:PATTern:TIMeout:MODE
<TimeoutMode>
Defines the condition for the timeout.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<TimeoutMode>
HIGH | LOW
HIGH
The result stays high.
LOW
The result stays low.
*RST:
HIGH
Usage:
Asynchronous command
TRIGger<m>:PATTern:TIMeout[:TIME]
<Time>
Defines how long the result of the pattern condition must be true or false.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Time>
Range:
100E-12 to 10000
Increment: 100E-9
*RST:
100E-9
Default unit: s
Usage:
Asynchronous command
TRIGger<m>:PATTern:WIDTh:RANGe
<WidthRangeMode>
Defines how the range of a pulse width is defined for keeping up the true result of the
pattern condition. The width and delta are specified using
TRIGger<m>:PATTern:WIDTh:DELTa
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<WidthRangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers on pulses inside a given range. The range is defined by
the width
±
delta.
OUTSide
Triggers on pulses outside a given range. The range is defined
by the width
±
delta.
Trigger