Instrument tour
R&S
®
RTO6
37
Getting Started 1801.6670.02 ─ 03
[Intensity]
Adjusts the intensity of the waveforms on the screen, or the background transpar-
ency of dialog boxes, or the transparency of result boxes.
Press the [Intensity] key and turn the [Multiuse] knob. Press the [Intensity] key
again to toggle between the three settings. The controlled parameter and its value
are shown in the input box in the upper right corner of the screen.
[Cursor]
Starts a cursor measurement: displays vertical and horizontal cursors in the
active diagram, and the results of the cursor measurement.
Cursors are markers which are placed at points of interest on a waveform. The
instrument measures the cursor positions and delta values between parallel cur-
sors.
If you press the key while a cursor measurement is enabled, the "Cursors" dialog
box opens.
In the "Cursors" dialog box, you can:
●
Configure up to 4 cursor sets
●
Define labels of the cursors
●
Connect the cursor to the waveform and couple the cursors
[Measure]
Starts the default automatic measurement for the active waveform and displays
the measurement results.
If you press the [Measure] key while a measurement is enabled, the "Measure-
ments" dialog box is displayed, where you can:
●
Configure various measurements and enable statistics
●
Configure gated measurement
●
Configure long term and histogram measurements
●
Configure actions to be executed if specified limits are exceeded
[History]
The sample memory contains several stored acquisitions before the current one,
which is shown in the display. Press the key to open the quick access "History"
dialog box, where you can view the stored acquisitions and use them for further
analysis. Press the key again to open the main "History" dialog box with more set-
tings and information.
The key is illuminated as long as a history acquisition or replay is displayed.