Remote Commands Reference
R&S
®
RTC1000
207
User Manual 1335.7352.02 ─ 02
TRIGger:A:PATTern:WIDTh[:WIDTh]
<PatternWidth>
For the ranges WITHin and OUTSide (defined using
), the <PatternWidth> defines the center of a range which is defined by the limits
±
<Delta> (set with
For the ranges SHORter and LONGer, the pattern width defines the maximum and
minimum values, respectively.
Parameters:
<PatternWidth>
Center value, maximum value or minimum value depending on
the defined range type.
Range:
1.600E-08 to 3.01
*RST:
3.9999999E-04
Default unit: s
TRIGger:A:PATTern:WIDTh:DELTa
<PatternDelta>
Defines a range around the pattern width value specified using
.
Parameters:
<PatternDelta>
±Δ
t ("Variation")
Range:
8.00E-09 to maximum value, which depends on
the defined pulse width
*RST:
1.5000000E-04
TRIGger:A:PATTern:WIDTh:RANGe
<PatternRange>
Defines a range around the width value specified using
.
Parameters:
<PatternRange>
WITHin | OUTSide | SHORter | LONGer
WITHin | OUTSide
Triggers on pulses inside or outside a range defined by time ±
delta. The time is specified with
, the range around is defined with
SHORter | LONGer
Triggers on pulses shorter or longer than a time set with
TRIGger:A:PATTern:WIDTh[:WIDTh]
*RST:
LONGer
Trigger
Содержание RTC1000
Страница 1: ...R S RTC1000 Digital Oscilloscope User Manual User Manual 1335 7352 02 02 S 2...
Страница 57: ...Trigger R S RTC1000 57 User Manual 1335 7352 02 02 Figure 5 3 External trigger signal External Trigger Input...
Страница 225: ...Remote Commands Reference R S RTC1000 225 User Manual 1335 7352 02 02 Suffix m 1 6 Selects the measurement Measurements...