User Manual 1335.7898.02 ─ 08
TRIGger:A:CAN:TYPE
<TriggerType>
Specifies the trigger mode for CAN.
Parameters:
<TriggerType>
STOFrame | EOFrame | ID | IDDT | FTYPe | ERRCondition
STOFrame
Start of frame
EOFrame
End of frame
ID
Sets the trigger to a specific message "Identifier" or an "Identi-
fier" range.
Specify the identifier with
.
IDDT
Sets the trigger to a combination of "Identifier and Data" condi-
tion. The instrument triggers at the end of the last byte of the
specified data pattern.
Specify the "Identifier" (see
ID
), and the "Data" with
and
FTYPe
Triggers on a specified "Frame".
Specify the frame type with
.
ERRCondition
Identifies various errors in the frame.
Specify the "Error" with
,
.
*RST:
STOF
TRIGger:A:CAN:FTYPe
<FrameType>
Specifies the frame type to be triggered on if
FTYPe
.
Parameters:
<FrameType>
DATA | REMote | ERRor | OVERload | ANY
*RST:
ERR
TRIGger:A:CAN:ITYPe
<IdentifierType>
Selects the length of the identifier: 11 bit for CAN base frames, or 29 bits for CAN
extended frames.
is set to
ID
,
IDDT
, or
FTYPe
(data
and remote frames).
Serial bus analysis