Remote control commands
R&S
®
RTA4000
643
User Manual 1335.7898.02 ─ 08
Parameters:
<DataLength>
Range:
0 to 8
Increment: 1
*RST:
1
Default unit: Byte
TRIGger:A:CAN:DATA
<Data>
Defines the data pattern. The number of bytes in the data pattern is defined with
.
is set to
IDDT
.
Parameters:
<Data>
String containing binary pattern with max. 64 bit. Characters 0,
1, and X are allowed. Make sure to enter complete bytes.
TRIGger:A:CAN:ACKerror
<AcknowledgeError>
Triggers on acknowledgement errors. An acknowledgement error occurs when the
transmitter does not receive an acknowledgment - a dominant bit during the Ack Slot.
is set to
ERRCondition
.
Parameters:
<AcknowledgeError> ON | OFF
*RST:
OFF
TRIGger:A:CAN:BITSterror
<BitStuffingError>
Triggers on bit stuffing errors.
See also:
is set to
ERRCondition
.
Parameters:
<BitStuffingError>
ON | OFF
*RST:
ON
TRIGger:A:CAN:CRCerror
<CRCerror>
Triggers on errors in the Cyclic Redundancy Check.
is set to
ERRCondition
.
Parameters:
<CRCerror>
ON | OFF
*RST:
OFF
Serial bus analysis