Troubleshooting
R&S
®
RT-ZC31
38
User Manual 1801.4949.02 ─ 01
Table 4-1: LED errors
Error state
Meaning
Protection mode
A temperature anomaly has been
detected and the device has entered
protection mode.
Remove the device from the conductor
being measured immediately.
Allow the device to cool under condi-
tions of no input and then press the
DEMAG/AUTO ZERO key. The device
returns to its state when it was turned
on.
Start measurement after performing
demagnetization and zero adjustment
again.
We recommended you to calibrate the
device since internal components may
have been subject to stress.
Overload
The input current exceeding the speci-
fied level of each current range is detec-
ted.
If you use the 30 A range, immediately
remove the sensor from the conductor
being measured.
If you use the 0.5 A range or 5 A range,
switch to a higher current range.
The instrument may not be able to prop-
erly detect overload states immediately
after the current range has changed.
The currents for which an overload
state can be detected are DC and sine
waves with frequencies of 45 Hz to 66
Hz. The device is unable to detect the
currents listed below as an overload
state.
●
Currents that exceed the defined
level on a momentary basis
●
High-frequency currents that
exceed the defined level
Although external magnetic fields may
cause the OVERLOAD LED to blink
while the upper jaw is retracted, this
does not indicate an issue with the
device
LED Display Errors