Measuring Procedure
R&S
®
RT-ZC31
31
User Manual 1801.4949.02 ─ 01
The DEMAG / AUTO ZERO LED lights up, and automatic zero-adjustment is
performed alone. After the completion, the DEMAG / AUTO ZERO LED goes
out.
6. Measure a current.
3.6
Protection Mode
To protect the device against self-generated heat, it enters protection mode when
the temperature of the junction box exceeds a specified level.
In protection mode, the device cannot correctly measure any current. Moreover,
you cannot switch the current ranges. If the device has entered protection mode,
recalibrate it because internal components may have been subjected to thermal
stress.
To exit the protection mode
1. Pull the opening lever toward you. Remove the device from the conductor that
you measure.
2. Slide the opening lever toward the lower jaw until the JAW UNLOCKED indi-
cator is hidden.