Remote Control Commands
R&S
®
NRX
378
User Manual 1178.5566.02 ─ 07
Suffix:
<undef>
.
1 to n
No suffix required.
Parameters:
<auto>
ON | OFF | ONCE | LFR | UFR
13.11
Running Selftests
Used for testing the connected power sensors and the R&S NRX.
Further information:
●
.......................................................................................... 378
CALibration<Sensor>:TEST:DEViation?
........................................................................... 378
CALibration<Sensor>:TEST:REFerence?
......................................................................... 379
...............................................................................................379
................................................................................................380
.................................................................................. 380
CALibration<Sensor>:TEST?
Requires a thermal power sensor, equipped with an additional test heater.
Queries the power difference in W when measuring with the external heater enabled
and disabled.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
CALibration<Sensor>:TEST:DEViation?
Requires a thermal power sensor, equipped with an additional test heater.
Queries the relative deviation of the current power difference from the reference value
stored in the calibration data set. The relative deviation is calculated as follows:
<relative
deviation>
= (<current
power
difference>
/ <reference
value>) - 1
with
<current power difference>:
CALibration<Sensor>:TEST:REFerence?
Running Selftests