Remote Control Commands
R&S
®
NRX
316
User Manual 1178.5566.02 ─ 05
Parameters:
<value>
E1 | E2 | E4 | E8 | E16 | E32 | E64 | E128 | E256
*RST:
E4
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TCONtrol:AUTO
<state>
Effective for continuous average, burst average, pulse analysis, timeslot, statistics
measurements.
Enables or disables the automatic termination control. See also
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TCONtrol[:
ENUM]
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
*RST:
ON
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TCONtrol[:ENUM]
<mode>
Effective for continuous average, burst average, pulse analysis, timeslot, statistics
measurements.
Defines how the measurement results are output. This is called termination control.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<mode>
MOVing | REPeat
MOVing
Outputs intermediate values to facilitate early detection of
changes in the measured quantity. In the settled state, that
means when the number of measurements specified by the
average count has been performed, a moving average is output.
Configuring Sensors