Remote Control Commands
R&S
®
NRX
173
User Manual 1178.5566.02 ─ 05
ABORt<Measurement>
Immediately interrupts the current measurement. If the measurement has been started
as a single measurement (
INITiate<Measurement>[:IMMediate]
sensor goes into the idle state. However, if a continuous measurement is in progress
(
INITiate<Measurement>:CONTinuous
ON
), the trigger system of the power sen-
sor enters the waiting for trigger state, and if the trigger condition is met, a new mea-
surement is immediately started.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Usage:
Event
ABORt<undef>:ALL
Applies to all connected power sensors. See
Suffix:
<undef>
.
1 to n
No suffix required.
Usage:
Event
INITiate<Measurement>:CONTinuous
<state>
Enables or disables the continuous measurement mode. In continuous measurement
mode, the power sensor does not go into the idle state after a measurement has been
completed, but immediately executes another measurement cycle.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<state>
ON
Measurements are performed continuously. If a measurement is
completed, the power sensor does not return to the idle state but
enters the waiting for trigger state again.
OFF
Ends the continuous measurement mode, and sets the power
sensor to the idle state.
*RST:
0
Starting and Ending a Measurement