TRIGger:DELay
<delay>
Sets the delay between the trigger event and the beginning of the actual measurement
(integration).
Parameters:
<delay>
Range:
-5.0 to 10.0
*RST:
0.0
Default unit: s
Manual operation:
See
TRIGger:DELay:AUTO
<state>
If
ON
is set, no measurement is started until the power sensor
has settled. For this purpose, the delay value is automatically determined.
If a longer period is set using
, the automatically determined delay is
ignored.
Parameters:
<state>
*RST:
OFF
TRIGger:DTIMe
<dropout_time>
Sets the dropout time for the internal trigger source. During this time, the signal power
must exceed (negative trigger slope) or undercut (positive trigger slope) the level
defined by the trigger level and trigger hysteresis. At least, this time must elapse before
triggering can occur again.
See
Chapter 9.5.2.4, "Dropout Time"
Parameters:
<dropout_time>
Range:
0.00 to 10.00
*RST:
0.00
Default unit: s
Manual operation:
See
TRIGger:EXTernal<2...2>:IMPedance
<impedance>
Effective only if
EXTernal2
is set.
Sets termination resistance of the second external trigger input. Choose the setting
that fits the impedance of the trigger source to minimize reflections on the trigger sig-
nals.