Remote Control - Commands
R&S
®
NRP2
258
User Manual 1173.9157.02 ─ 03
OUTPut:TTL:FEED
<calculate_block>
Sets the calculate block whose limit handling controls the TTL voltage. If the measured
value of this calculate block violates a limit and limit monitoring is active, the voltage
specified in
(first line) is output.
Parameters:
<calculate_block>
string
Range:
"CALC1:LIM","CALC2:LIM","CALC3:LIM","CALC4:L
IM"
*RST:
"CALC1:LIM"
OUTPut:TTL:HVOLtage
<high_voltage>
Sets the high voltage for the TTL output.
Parameters:
<high_voltage>
float_value
Range:
0 V to 3.3 V
*RST:
3.3 V
Default unit: V
OUTPut:TTL:LVOLtage
<low_voltage>
Sets the low voltage for the TTL output.
Note:
The value for the lower TTL level may be greater than the value for the upper TTL
level.
Parameters:
<low_voltage>
float_value
Range:
0 V to 3.3 V
*RST:
0 V
Default unit: V
OUTPut:TTL[:STATe]
<state>
Activates/deactivates the TTL operating mode of the OUT1 connector (rear panel). The
magnitude of the two possible output voltages can range from 0 V to 3.3 V
(
OUTP:TTL:HVOL
and
OUTP:TTL:LVOL
to determine which voltage is available
at the output.
Table 6-14: Voltage at TTL output
OUTP:TTL:ACT LOW
(OUTP:TTL:FAIL HIGH)
OUTP:TTL:ACT HIGH
(OUTP:TTL:FAIL LOW)
Measured value violates limit
OUTP:TTL:LVOL
OUTP:TTL:HVOL
Measured value is within limit
OUTP:TTL:HVOL
OUTP:TTL:LVOL
OUTPut
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...