User Manual 1177.6017.02 ─ 08
The complex reflection coefficient
Γ
sensor
of the sensor, which is also required for the
correction, is prestored in the calibration data memory for many frequencies.
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Sensor
Source
Figure 9-2: Correction of interactions between the power sensor and the signal source
If the gamma correction is performed in combination with an S-parameter correction
(
[SENSe<Sensor>:]CORRection:SPDevice:STATe
ON
), the following is consid-
ered:
●
Interaction of the signal source with the S-parameter device
●
Input of the power sensor, depending on the transmission, expressed by the term
s
12
s
21
The interaction between the complex reflection coefficient
Γ
sensor
of the power sensor
and the reflection of port
2 is expressed by
s
22
rection is enabled, this interaction is always considered, regardless whether gamma
correction is performed or not.
Sensor
Source
S
21
S
12
S
22
S
11
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 9-3: Correction of interactions between the power sensor, the signal source, and the S-param-
eter device
Remote commands:
[SENSe<Sensor>:]SGAMma:CORRection:STATe
................................................................93
[SENSe<Sensor>:]SGAMma:MAGNitude
........................................................................... 94
................................................................................. 94
[SENSe<Sensor>:]SGAMma:CORRection:STATe
<state>
Enables or disables the use of the complex reflection coefficient to correct the interac-
tions of power sensor and signal source.
Parameters:
<state>
*RST:
OFF
Manual operation:
See
Configuring basic measurement parameters