Remote Control
R&S
®
NGE100B
89
User Manual 5601.1343.02 ─ 04
Return values:
<Power>
<Numeric_value>
Return the measured power value.
Default unit: W
Example:
MEAS:POW?
-> 3.00E+00
Return the measured power value of the selected channel.
Usage:
Query only
MEASure[:SCALar][:VOLTage][:DC]?
This command queries the measured voltage value of the previous selected channel.
Return values:
<Voltage>
<numeric_value>
Return the measured voltage.
Default unit: V
Example:
MEAS?
-> 1.000E+00
Usage:
Query only
6.5.11
Voltage and Current Configuration Commands
APPLy
<arg0>
This command defines or queries the voltage and current value of the selected chan-
nel.
Parameters:
<arg0>
{<Voltage> | DEF | MIN | MAX} [, {<Current> | DEF | MIN | MAX}]
<Voltage>
Set voltage in the range of 0 V to 32.20 V.
DEF
Set voltage to default.
MIN
Set voltage to minimum .
MAX
Set voltage to maximum.
<Current>
Set current in the range of 3 mA to 3.025 A.
DEF
Set current to default.
MIN
Set currernt to minimum.
MAX
Set current to maximum.
Remote Control - Commands