Instrument Functions
R&S
®
NGE100B
41
User Manual 5601.1343.02 ─ 04
2. Press any of the memory keys (
M1
,
M2
,
M3
,
M4
,
M5
) to store the instrument set-
tings.
The R&S NGE100B stores the instruments settings in the selected memory loca-
tion.
Recall instrument settings
1. Press
Store Recall
key to recall instrument settings.
The R&S NGE100B displays a message in the status bar "Recall device settings:
select any [Mx] key".
2. Press any of the memory keys (
M1
,
M2
,
M3
,
M4
,
M5
) to recall the instrument set-
tings.
The R&S NGE100B recalls the instruments settings from the selected memory
location.
5.5
Advanced Functions
The following chapters provide information about the menu items found in the
Menu
key.
1. Press
Menu
key.
The R&S NGE100B displays the main menu of the instrument.
See
2. Select the desired menu item.
The R&S NGE100B displays the respective menu item dialog.
5.5.1
Protection
The R&S NGE100B provides individual overvoltage protection (OVP) and overpower
protection (OPP) settings for each channel. With both function activated, the R&S
NGE100B protects itself and the load in case of voltage or power overloads.
When protection is tripped, a red flashing OVP or OPP indicator displays on the
respective channel. See
.
Advanced Functions