Remote Commands
R&S
®
FSW
870
User Manual 1173.9411.02 ─ 43
<gap>
1 | 2
Gap (CACLR) channel number
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:LIM:ACP:GAP1:ACLR:REL:STAT ON
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:AUTO][:CACLr][:RELative]
<Limit>, <UpperLimit>
This command defines the relative limit of the specified gap (CACLR) channel. The ref-
erence value for the relative limit is the measured channel power.
If you define both an absolute limit and a relative limit, the R&S
FSW uses the lower
value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<gap>
1 | 2
Gap (CACLR) channel number
Parameters:
<Limit>
Defines the relative limit of the specified gap channel in dB.
Default unit: DB
<UpperLimit>
Optional: Defines the relative upper limit of the specified gap
channel.
Default unit: dB
Example:
CALC:LIM:ACP:GAP1:REL 3dB,0
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:AUTO][:CACLr][:RELative]:STATe
<State>
This command turns the relative limit check for the specified gap (CACLR) channel on
and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<gap>
1 | 2
Gap (CACLR) channel number
Parameters:
<State>
ON | OFF | 0 | 1
Configuring and Performing Measurements
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...