Gate settings define one or more extracts of the signal to be measured.
Gating is not available for measurements on I/Q-based data.
Gated Trigger
Switches gated triggering on or off.
If the gate is switched on, a gate signal applied to one of the TRIGGER INPUT connec-
tors or the internal IF power detector controls the sweep.
Remote command:
Gate Mode
Sets the gate mode.
For more information see
Chapter 8.6.2.1, "Gated Measurements"
"Edge"
The trigger event for the gate to open is the detection of the signal
edge.
After the gate signal has been detected, the gate remains open until
the gate length is over.
Trigger and Gate Configuration