Measurements and Results
R&S
®
FSW
195
User Manual 1173.9411.02 ─ 43
How to Perform a Standard Channel Power Measurement
..................................195
.................................................................................195
How to Configure an MSR ACLR Measurement
...................................................196
How to Manage User-Defined Configurations
.......................................................198
How to Compare the Tx Channel Power in Successive Measurements
7.2.6.1
How to Perform a Standard Channel Power Measurement
Performing a channel power or ACLR measurement according to common standards is
a very easy and straightforward task with the R&S
FSW.
1. Press the [MEAS] key or select "Select Measurement" in the "Overview".
2. Select "Channel Power ACLR".
The measurement is started immediately with the default settings.
3. Select the "CP / ACLR Standard" softkey.
4. Select a standard from the list.
The measurement is restarted with the predefined settings for the selected stan-
dard.
5. If necessary, edit the settings for your specific measurement as described in
ter 7.2.6.2, "How to Set Up the Channels"
on page 195, or load a user-defined
"To load a user-defined configuration"
7.2.6.2
How to Set Up the Channels
Channel definition is the basis for measuring power levels in certain frequency ranges.
Usually, the power levels in one or more carrier (Tx) channels and possibly the adja-
cent channels are of interest. Up to 18 carrier channels and up to 12 adjacent channels
can be defined.
When a measurement standard is selected, all settings including the channel band-
widths and channel spacings are set according to the selected standard. Select a stan-
dard in the "Ch Power" menu or the "ACLR Setup" dialog box. You can adjust the set-
tings afterwards.
Channel setup consists of the following settings:
●
The number of transmission (Tx) and adjacent channels
●
The bandwidth of each channel
●
For multicarrier ACLR measurements: which Tx channel is used as a reference
●
The spacing between the individual channels
●
Optionally: the names of the channels displayed in the diagram and result table
●
Optionally: the influence of individual channels on the total measurement result
("Weighting Filter")
●
Optionally: limits for a limit check on the measured power levels
Channel Power and Adjacent-Channel Power (ACLR) Measurement
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...