Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1029
User Manual 1178.8520.02 ─ 01
Parameters:
<Resolution>
0.001 | 0.01 | 0.1 | 1 | 10 | 100 | 1000 | 10000 Hz
*RST:
0.1 Hz
Default unit: HZ
Example:
CALC:MARK:COUN:RES 1kHz
Sets the resolution of the frequency counter to 1 kHz.
Manual operation:
See
14.8.3.14
Marker Demodulation
The following commands control the demodulation of AM and FM signals at the marker
position.
Useful commands for marker demodulation described elsewhere:
●
Remote commands exclusive to marker demodulation:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
..................................1029
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff
........................................1030
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect
..........................................1030
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe]
......................................... 1030
...................................................................................1031
................................................................................ 1031
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
<State>
This command turns continuous demodulation of the signal at the marker position in
the frequency domain on and off.
In the time domain continuous demodulation is always on.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC2:MARK3:FUNC:DEM:CONT ON
Switches on the continuous demodulation.
Manual operation:
See
Analyzing Measurements (Basics)