Configuration
R&S
®
FSV/A3000 I/Q
Analyzer
54
User Manual 1178.8536.02 ─ 01
For details, see
Chapter 4.4, "Basics on Input from I/Q Data Files"
I/Q Input File State
Enables input from the selected I/Q input file.
If enabled, the application performs measurements on the data from this file. Thus,
most measurement settings related to data acquisition (attenuation, center frequency,
measurement bandwidth, sample rate) cannot be changed. The measurement time
can only be decreased, to perform measurements on an extract of the available data
only.
Note:
Even when the file input is disabled, the input file remains selected and can be
enabled again quickly by changing the state.
Remote command:
Select I/Q data file
Opens a file selection dialog box to select an input file that contains I/Q data.
The I/Q data must have a specific format (
.iq.tar
) as described in
The default storage location for I/Q data files is
C:\R_S\INSTR\USER
.
Remote command:
5.3.1.3
Probe Settings
Access
: [INPUT / OUTPUT] > "Input Source Config" > "Probes"
Data input for the measurement can be provided by probes if the optional R&S RT-ZA9
adapter is used.
Data Input and Output Settings