Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
143
User Manual 1178.8536.02 ─ 01
INPut<ip>:FILE:PATH
<FileName>[, <AnalysisBW>]
This command selects the I/Q data file to be used as input for further measurements.
The I/Q data must have a specific format as described in
Chapter C, "I/Q Data File For-
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<FileName>
String containing the path and name of the source file. The file
extension is
*.iq.tar
.
<AnalysisBW>
Optionally: The analysis bandwidth to be used by the measure-
ment. The bandwidth must be smaller than or equal to the band-
width of the data that was stored in the file.
Default unit: HZ
Example:
INP:FILE:PATH 'C:\R_S\Instr\user\data.iq.tar'
Uses I/Q data from the specified file as input.
Manual operation:
See
8.4.1.3
Setting up Probes
Modular probes can be connected to the RF input connector of the R&S
FSV/A.
For details see
[SENSe:]PROBe<pb>:ID:PARTnumber?
.......................................................................... 143
[SENSe:]PROBe<pb>:ID:SRNumber?
..............................................................................144
[SENSe:]PROBe<pb>:SETup:ATTRatio
............................................................................144
[SENSe:]PROBe<pb>:SETup:CMOFfset
.......................................................................... 144
[SENSe:]PROBe<pb>:SETup:DMOFfset
.......................................................................... 145
............................................................................... 145
..............................................................................146
[SENSe:]PROBe<pb>:SETup:NMOFfset
.......................................................................... 146
............................................................................. 146
[SENSe:]PROBe<pb>:SETup:PMOFfset
.......................................................................... 147
[SENSe:]PROBe<pb>:SETup:STATe?
..............................................................................147
...............................................................................148
[SENSe:]PROBe<pb>:ID:PARTnumber?
Queries the R&S part number of the probe.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Return values:
<PartNumber>
Configuring I/Q Analyzer Measurements