Basics on I/Q Data Acquisition and Processing
R&S
®
FSV/A3000 I/Q
Analyzer
20
User Manual 1178.8536.02 ─ 01
Maximum record length for RF input
The maximum record length, that is, the maximum number of samples that can be cap-
tured, is 100
MSamples (with option B114: 800
MSamples).
shows the maximum usable I/Q bandwidths depending on the output
sample rates.
Figure 4-2: Relationship between maximum usable I/Q bandwidth and output sample rate with and
without bandwidth extensions
4.1.1.3
R&S
FSV/A Without Additional Bandwidth Extension Options
Sample rate: 100 Hz - 10
GHz
Maximum I/Q bandwidth: 10 MHz
Table 4-1: Maximum I/Q bandwidth
Sample rate
Maximum I/Q bandwidth
100 Hz to 10 MHz
Proportional up to maximum 10 MHz
10 MHz to 10
GHz
10 MHz
4.1.1.4
R&S
FSV/A with I/Q Bandwidth Extension Option B40 or U40
Sample rate: 100 Hz - 10
GHz
Maximum bandwidth: 40 MHz
Sample rate
Maximum I/Q bandwidth
100 Hz to 50 MHz
Proportional up to maximum 40 MHz
50 MHz to 10
GHz
40 MHz