Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
179
User Manual 1178.8536.02 ─ 01
For R&S
FSV/A4, 7, 13, and 30 models: 15 dB and 30 dB
R&S
FSV/A44 or higher: 30 dB
All other values are rounded to the nearest of these two.
Default unit: DB
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Manual operation:
See
8.4.2.4
Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
................................................................... 179
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
............................................... 179
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
..............................180
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision
.................................................... 180
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
................................................... 180
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
........................................181
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
<Range>
This command defines the display range of the y-axis (for all traces).
Note that the command works only for a logarithmic scaling. You can select the scaling
with
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Range>
Range:
1 dB to 200 dB
*RST:
100 dB
Default unit: HZ
Example:
DISP:TRAC:Y 110dB
Manual operation:
See
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again (for all
traces).
Suffix:
<n>
.
<t>
irrelevant
Configuring I/Q Analyzer Measurements