Configuration
R&S
®
FSV/A3000 I/Q
Analyzer
70
User Manual 1178.8536.02 ─ 01
Scaling
Defines the scaling method for the y-axis.
"Logarithmic"
Logarithmic scaling (only available for logarithmic units - dB..., and A,
V, Watt)
"Linear with
Unit"
Linear scaling in the unit of the measured signal
"Linear Per-
cent"
Linear scaling in percentages from 0 to 100
"Absolute"
The labeling of the level lines refers to the absolute value of the refer-
ence level (not available for "Linear Percent" )
"Relative"
The scaling is in dB, relative to the reference level (only available for
logarithmic units - dB...). The upper line of the grid (reference level) is
always at 0 dB.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
Y-Axis Max
Defines the maximum value of the y-axis in the currently selected diagram in either
direction (in Volts). Thus, the y-axis scale starts at -<Y-Axis Max> and ends at +<Y-Axis
Max>.
This command is only available if the evaluation mode for the I/Q Analyzer is set to
"I/Q-Vector" or "Real/Imag (I/Q)" .
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
5.5
Frequency Settings
Access
: "Overview" > "Frequency"
Frequency Settings