Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
173
User Manual 1178.8536.02 ─ 01
................................................................................................ 173
...................................................................................174
.....................................................................................177
8.4.2.1
Amplitude Settings
Useful commands for amplitude configuration described elsewhere:
●
Remote commands exclusive to amplitude configuration:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
........................................................... 173
........................................................................................... 173
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
........................................................174
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
........................................... 174
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Suffix:
<n>
.
<m>
Example:
CALC:MARK2:FUNC:REF
Sets the reference level to the level of marker 2.
Manual operation:
See
" Reference Level = Marker Level "
CALCulate<n>:UNIT:POWer
<Unit>
UNIT<n>:POWer
<Unit>
This command selects the unit of the y-axis.
The unit applies to all power-based measurement windows with absolute values.
Suffix:
<n>
.
irrelevant
Parameters:
<Unit>
DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT |
DBUA | AMPere
*RST:
dBm
Example:
UNIT:POW DBM
Sets the power unit to dBm.
Configuring I/Q Analyzer Measurements