Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
77
Operating Manual 1176.7578.02 ─ 09
The scaling of the capture buffer is
●
relative to the current reference level if you are using the RF input and
●
relative to the full scale level if you are using the I/Q input
Available for source types:
●
Capture Buffer
●
Meas & Ref Signal
●
Error Vector
Figure 3-7: Result display "Real/Imag (I/Q)" in normal mode
SCPI commands:
CALC:FEED 'XTIM:DDEM:MEAS'
to define the required source type (see
CALC:FORM RIMag
to define the result type (see
TRAC:DATA TRACE1
to query the trace results (see
3.1.1.8
Eye Diagram Real (I)
The eye pattern of the inphase (I) channel; the x-axis value range is from -1 to +1 sym-
bols (MSK: -2 to +2)
Available for source types:
●
Meas & Ref Signal
Display lines are available in eye diagrams which allow you to determine the size of the
eye, see also
Chapter 3.3.9, "Measuring the Size of an Eye"
Measurement Result Display