9. Select the "Limit Checking On" option, or press the "ModAcc Limits On" softkey in
the "Limits" menu.
The limit check is performed immediately on the current modulation accuracy mea-
surement results and for all subsequent measurements until it is deactivated. The
results of the limit check are indicated by red or green values in the result sum-
mary.
3.3.9 Measuring the Size of an Eye
Display lines are available for eye diagrams, which allow you to determine the size of
the eye.
1. Configure a measurement with "Meas & Ref" as the "Source" and an eye diagram
as the "Result type" (in the
dialog box).
2. Press the LINES key on the front panel.
3. Select the "Vertical Line Absolute" softkey.
4. Move the line to the left edge of the eye.
5. Select the "Vertical Line Rel" softkey.
6. Move the line to the right edge of the eye.
The relative position of the line is indicated in the diagram, which corresponds to
the width of the eye (see
7. Select the "Horizontal Line Absolute" softkey.
8. Move the line to the top of the eye.
9. Select the "Horizontal Line Rel" softkey.
10. Move the line to the bottom of the eye.
The relative position of the line is indicated in the diagram, which corresponds to
the height of the eye (see
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...