Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
201
Operating Manual 1176.7578.02 ─ 09
Figure 3-35: Size of an eye, calculated using display lines
3.4 Further Information
3.4.1
3.4.2
3.4.1 Trace Mode Overview
The traces can be activated individually for a measurement or frozen after completion
of a measurement. Traces that are not activated are hidden. Each time the trace mode
is changed, the selected trace memory is cleared.
The R&S
FSVA/FSV provides the following different trace modes:
Clear Write
Overwrite mode: the trace is overwritten by each sweep. This is the default setting.
Remote command:
DISP:TRAC:MODE WRIT
, see
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Max Hold
The maximum value is determined over several sweeps and displayed. The
R&S
FSVA/FSV saves the sweep result in the trace memory only if the new value is
greater than the previous one.