Setting parameters:
<RefDevCompensation>
ON | OFF
ON
Scales the reference signal to the current deviation of the mea-
surement signal
OFF
Uses the nominal deviation you have set for the reference signal
*RST:
ON
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:NORMalize:IQIMbalance
<CompIQImbalance>
This command switches the compensation of the IQ imbalance on or off.
Setting parameters:
<CompIQImbalance> ON | OFF
*RST:
OFF
Example:
DDEM:NORM:IQIM OFF
Switches the compensation off.
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:NORMalize:IQOFfset
<CompIQOffset>
This command switches the compensation of the IQ offset on or off.
Setting parameters:
<CompIQOffset>
ON | OFF
*RST:
ON
Example:
DDEM:NORM:IQOF OFF
Switches the compensation off.
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:NORMalize[:VALue]
<Normalize>
This command switches the compensation of the I/Q offset and the compensation of
amplitude droop on or off. When queried, the command returns 1 if both are ON and 0
if both are off. Otherwise, an error is returned.
The command is kept because of compatibility to the R&S FSQ and won't be suppor-
ted in later versions. Instead, use the new command (
SENSe subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...