Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
138
Operating Manual 1176.7578.02 ─ 09
Firmware Update ← EXIQ
If a firmware update for the R&S EX-IQ-BOX is delivered with the R&S
FSVA/FSV firm-
ware, this function is available. In this case, when you select the softkey, the firmware
update is performed.
R&S Support ← EXIQ
Stores useful information for troubleshooting in case of errors.
This data is stored in the
C:\R_S\Instr\user\Support
directory on the instru-
ment.
The
SupportSave.dfl
file contains the instrument settings and input data and can
be loaded to the instrument again for inspection later. (Remember to set the sweep
mode to "Single Sweep" beforehand, as "Continuous Sweep" would immediately over-
write the loaded input data.)
If you contact the Rohde&Schwarz support to get help for a certain problem, send
these files to the support in order to identify and solve the problem faster.
DigIConf ← EXIQ
Starts the optional R&S DigIConf application. This softkey is only available if the
optional software is installed.
To return to the R&S
FSVA/FSV application, press any key on the front panel. The
application is displayed with the "EXIQ" menu, regardless of which key was pressed.
For details on the R&S DigIConf application, see the "R&S®EX-IQ-BOX Digital Inter-
face Module R&S®DigIConf Software Operating Manual".
Note:
If you close the R&S DigIConf window using the "Close" icon, the window is
minimized, not closed.
If you select the "File > Exit" menu item in the R&S DigIConf window, the application is
closed. Note that in this case the settings are lost and the EX-IQ-BOX functionality is
no longer available until you restart the application using the "DigIConf" softkey in the
R&S
FSVA/FSV once again.
Remote command:
Remote commands for the R&S DigIConf software always begin with
SOURce:EBOX
.
Such commands are passed on from the R&S
FSVA/FSV to the R&S DigIConf auto-
matically which then configures the R&S EX-IQ-BOX via the USB connection.
All remote commands available for configuration via the R&S DigIConf software are
described in the "R&S®EX-IQ-BOX Digital Interface Module R&S®DigIConf Software
Operating Manual".
Example 1
:
SOURce:EBOX:*RST
SOURce:EBOX:*IDN?
Result:
"Rohde&Schwarz,DigIConf,02.05.436 Build 47"
Example 2
:
SOURce:EBOX:USER:CLOCk:REFerence:FREQuency 5MHZ
Defines the frequency value of the reference clock.
Input (AC/DC)
Toggles the RF input of the R&S
FSVA/FSV between AC and DC coupling.
Softkeys and Menu Overview for Vector Signal Analysis (R&S
FSV-K70)
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...