Troubleshooting
R&S
®
FSMR3000
694
User Manual 1179.0116.02 ─ 02
Table 13-3: Device-specific error messages
Error code
Error text in the case of queue poll
Error explanation
1052
Frontend LO is Unlocked
This message is displayed when the phase regulation of the local oscillator fails in the RF
front-end.
1060
Trigger-Block Gate Delay Error- gate length < Gate Delay
This message is displayed when the gate signal length is not sufficient for the pull-in delay
with a predefined gate delay.
1064
Tracking LO is Unlocked
This message is displayed when the phase regulation of the local oscillator fails on the
external generator module.
2028
Hardcopy not possible during measurement sequence
This message is displayed when a printout is started during scan sequences that cannot
be interrupted. Such sequences are for example:
●
Recording the system error correction data (alignment)
●
Instrument self-test
In such cases synchronization to the end of the scan sequence should be performed prior
to starting the printout.
2033
Printer Not Available
This message is displayed when the selected printer is not included in the list of available
output devices. A possible cause is that the required printer driver is missing or incorrectly
installed.
2034
CPU Temperature is too high
This message is displayed when the temperature of the processor exceeds 70 °C.
13.3
Troubleshooting Remote Operation
If problems arise during measurement in remote operation, try the following methods to
solve them.
Incompleted sequential commands - blocked remote channels
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish and the remote
channel to the R&S
FSMR3 is blocked for further commands. In this case, you must
interrupt processing on the remote channel in order to abort the measurement.
To regain control over a blocked remote channel
Usually, if you wait a minute for the VISA connection to detect the lost connection and
clear the control channel by itself, you can then re-establish the connection again. If
this fails, try the following:
1. Press "Local" on the front panel of the R&S
FSMR3000 to return to manual opera-
tion (if not disabled). Then re-establish the connection.
Troubleshooting Remote Operation