Remote Commands
R&S
®
FSMR3000
446
User Manual 1179.0116.02 ─ 02
OUTPut<up>:TRIGger<tp>:OTYPe
<OutputType>
This command selects the type of signal generated at the trigger output.
Suffix:
<up>
.
1..n
<tp>
Selects the trigger port to which the output is sent.
2 = trigger port 2 (rear panel)
Parameters:
<OutputType>
DEVice
Sends a trigger signal when the R&S
FSMR3 has triggered
internally.
TARMed
Sends a trigger signal when the trigger is armed and ready for
an external trigger event.
UDEFined
Sends a user defined trigger signal. For more information see
*RST:
DEVice
Manual operation:
See
OUTPut<up>:TRIGger<tp>:PULSe:IMMediate
This command generates a pulse at the trigger output.
Suffix:
<up>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (rear)
<tp>
1..n
Manual operation:
See
OUTPut<up>:TRIGger<tp>:PULSe:LENGth
<Length>
This command defines the length of the pulse generated at the trigger output.
Suffix:
<up>
.
1..n
<tp>
Selects the trigger port to which the output is sent.
2 = trigger port 2 (rear)
Parameters:
<Length>
Pulse length in seconds.
Default unit: S
Example:
OUTP:TRIG2:PULS:LENG 0.02
Manual operation:
See
Configuring the Measurement