Common Analysis and Display Functions
R&S
®
FSMR3000
185
User Manual 1179.0116.02 ─ 02
Storing and Recalling Limit Lines
Limit lines can be stored with the configuration settings so they can be recalled for
other measurements at a later time. Note, however, that any changes made to the limit
lines
after
storing the configuration file cannot be restored and will be overwritten by
the stored values when the configuration file is recalled. Always remember to store the
settings again after changing the limit line values.
After recalling measurement settings, the limit line values applied to the measurement
may be different to those displayed in the "Limit Lines" dialog box.
For more information see the R&S
FSMR3 User Manual, chapter "Data Management".
Limit Line Settings and Functions
Stored limit line settings
When storing and recalling limit line settings, consider the information provided in the
Data Management chapter of the R&S
FSMR3 User Manual.
.........................................................................................185
Limit Line Management
Access
: "Overview" > "Analysis" > "Lines" > "Limit Lines"
or
: [LINES] > "Line Config" > "Limit Lines"
For the limit line overview, the R&S
FSMR3 searches for all stored limit lines with the
file extension
.LIN
in the
limits
subfolder of the main installation folder. The over-
view allows you to determine which limit lines are available and can be used for the
current measurement.
For details on settings for individual lines see
Name
The name of the stored limit line.
Result Display Configuration