7.6.4.1
AF Evaluation
Access
: "Overview" > "Demod Settings" > "Scaling"
Or
: "Meas Setup" > "Demod" > "Scaling" tab
These settings are only available for AF evaluations.
Dev per Division/ dB per Division
................................................................................165
............................................................................................166
Dev per Division/ dB per Division
Defines the modulation depth or the phase deviation or frequency deviation per divi-
sion (logarithmic: 0.1 to 20 dB):
AM display:
0.0001 % to 1000 %
FM display:
1 Hz/div to 100 MHz/div
PM display:
0.0001 rad/div to 1000 rad/div
Note:
The value defined per division refers to the default display of 10 divisions on the
y-axis. If fewer divisions are displayed (e.g. because the window is reduced in height),
the range per division is increased in order to display the same result range in the
smaller window. In this case, the per division value does not correspond to the actual
display.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision