Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
78
User Manual 1179.0880.02 ─ 03
8.5.4
Frequency configuration
........................................................................................... 78
..................................................................................78
........................................................................................... 78
............................................................................................ 79
[SENSe:]FREQuency:CENTer
<Frequency>
This command defines the center frequency.
Parameters:
<Frequency>
The allowed range and f
max
is specified in the data sheet.
*RST:
fmax/2
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Manual operation:
See
[SENSe:]FREQuency:CENTer:STEP
<StepSize>
This command defines the center frequency step size.
Parameters:
<StepSize>
f
max
is specified in the data sheet.
Range:
1 to fMAX
*RST:
0.1 x span
Default unit: Hz
Example:
//Set the center frequency to 110 MHz.
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Manual operation:
See
[SENSe:]FREQuency:OFFSet
<Offset>
This command defines a frequency offset.
If this value is not 0
Hz, the application assumes that the input signal was frequency
shifted outside the application. All results of type "frequency" will be corrected for this
shift numerically by the application.
Parameters:
<Offset>
Range:
-1 THz to 1 THz
*RST:
0 Hz
Default unit: HZ
Configuration