R&S ESU
Instrument Functions
Tracking Generator – Option R&S FSU-B9
4.324
Operating Manual 1302.6163.12 - 03
Fig. 4.23 Normalized measurement, shifted with REF VALUE POSITION 50%
Remote command:
DISP:WIND:TRAC:Y:RPOS 10PCT
REF VALUE
The
REF VALUE
softkey activates the input of a value which is assigned to the ref-
erence line.
With default settings the reference line corresponds to a difference of 0 dB between
the currently measured trace and the reference trace. Setting the
REF VALUE
to a
different value helps to compensate for changes to the level conditions in the signal
path after the calibration data have been recorded. If e.g. after a source calibration a
10 dB attenuation is inserted into the signal path between DUT and R&S ESU input,
the measurement trace will be moved by 10 dB down. Entering a
REF VALUE
of –
10 dB will then result in the reference line for difference calculation being moved by
10 dB down as well. This means that the measured trace will be placed on it, as dis-
played in
REF VALUE
always refers to the active window.
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...