Remote Control
R&S
®
ESR
636
User Manual 1175.7068.02 ─ 12
INSTrument:DELete
<ChannelName>
This command deletes a spectrum display.
Also see
●
●
Chapter 11.17.16, "Usage of Four Spectrum Instances"
Parameters:
<ChannelName>
String containing the name of the channel you want to delete.
A channel must exist in order to be able delete it.
Example:
INST:DEL 'Spectrum 4'
Deletes the fourth spectrum display.
INSTrument[:SELect]
<Mode> | <ChannelName>
This command activates a new measurement channel with the defined channel type,
or selects an existing measurement channel with the specified name.
●
●
Chapter 11.17.16, "Usage of Four Spectrum Instances"
Parameters:
<Mode>
RECeiver
Receiver mode
RTIMe
Realtime mode
SANalyzer
Spectrum mode
Spectrum 2 | Spectrum 3 | Spectrum 4
Selects one of the additional spectrum displays. The first spec-
trum display is always active. You can select it with
SANalyzer
.
A spectrum display must exist in order to be able to select it.
*RST:
RECeiver
11.4
Remote Commands in Receiver Mode
The following remote commands configure and perform EMI measurements. They are
available in receiver mode only.
Measurements and Result Displays
......................................................................637
.........................................................................................665
................................................................................. 667
........................................................................................670
Controlling Inputs and Outputs
..............................................................................672
Remote Commands in Receiver Mode