WCDMA Multi Evaluation Measurement
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
671
User Manual 1173.9657.02 ─ 11
<Expected3dB>
Tolerance value for power step size 3 dB
Range:
0 dB to 5 dB
*RST:
1.5 dB
Default unit: dB
<Expected4to7dB>
Tolerance value for power step size 4 dB to 7 dB
Range:
0 dB to 5 dB
*RST:
2.0 dB
Default unit: dB
Example:
See
Firmware/Software:
V1.0.4.11
CONFigure:WCDMa:MEAS<i>:MEValuation:LIMit:PCONtrol:HSDPcch
<Enable>,
<DTXtoNACK>, <NACKtoCQI>, <CQItoDTX>[, <TestCase>]
Defines nominal power steps for the HS-DPCCH limit set. Measurements at maximum
UE power and below maximum UE power are supported. Separate values can be
defined for the boundaries DTX > (N)ACK, (N)ACK > CQI and CQI > DTX. Additionally
the limit check can be enabled or disabled.
See also
chapter 3.2.5.3, "Power Control Limits"
Parameters:
<Enable>
OFF | ON
disables | enables the limit check
*RST:
ON
<DTXtoNACK>
Range:
-10 dB to 10 dB
*RST:
6.14 dB
Default unit: dB
<NACKtoCQI>
Range:
-10 dB to 10 dB
*RST:
-1.38 dB
Default unit: dB
<CQItoDTX>
Range:
-10 dB to 10 dB
*RST:
-4.76 dB
Default unit: dB
<TestCase>
T0DB | T1DB
T0DB
: measurement below maximum UE power with TPC com-
mand = 0 dB
T1DB
: measurement at maximum UE power with TPC com-
mand = 1 dB
*RST:
T1DB
Example:
See
Firmware/Software:
V1.0.4.11
V3.2.60: added <TestCase>
Manual operation:
See
Command Reference
深圳德标仪器
135-1095-0799