WCDMA Multi Evaluation Measurement
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
601
User Manual 1173.9657.02 ─ 11
Minimum Trigger Gap
Defines a minimum duration of the power-down periods (gaps) between two triggered
power pulses. This setting is valid for an "(IF) Power" trigger source.
The trigger system is controlled by means of a timer which is reset to zero in the follow-
ing instances:
●
At the IF power-down ramp of each triggered or untriggered pulse, even though the
previous counter may not have elapsed yet. A power-down ramp is detected when
the signal power falls below the trigger threshold.
●
At the beginning of each measurement: The minimum gap defines the minimum
time between the start of the measurement and the first trigger event.
The trigger system is re-armed as soon as the timer has reached the specified mini-
mum gap.
This parameter can be used to prevent unwanted trigger events due to fast power var-
iations.
Remote command:
3.3.2.5
Limit Settings
The "Limits" in the "Multi Evaluation Configuration" dialog define upper limits for the
modulation, power and spectrum results including the spectrum emission mask.
For details see
chapter 3.2.5, "Limit Settings and Conformance Requirements"
Fig. 3-27: Limit settings
Limits
The limits can be configured via the remote commands described in the following sec-
tions:
●
chapter 3.5.3.9, "Limits (Modulation)"
GUI Reference
135-1095-0799