The current DAU configuration by the user is incomplete or wrong and could not be
applied. Check especially the IP address configuration.
●
54 (Filesystem Error)
:
The hard disk of the DAU is full or corrupt. Please execute a DAU selftest.
●
101 (Firmware Error)
:
Indicates a firmware or software error. If you encounter this error for the first time,
restart the instrument.
If the error occurs again, consider the following hints:
–
Firmware errors can often be repaired by restoring the factory default settings.
To restore these settings, restart your instrument and press the "Factory
Default" softkey during startup.
–
If a software package (update) has not been properly installed this is often indi-
cated in the "Setup" dialog, section "SW/HW-Equipment > Installed Software".
–
A software update correcting the error may be available. Updates are e.g. pro-
vided in the "CMW Customer Web" on GLORIS (registration required):
.
If you get firmware errors even with the properly installed latest software version,
please send a problem report including log files to Rohde & Schwarz.
●
102 (Unidentified Error)
:
Indicates an error not covered by other reliability values. For troubleshooting
please follow the steps described for "101 (Firmware Error)".
●
103 (Parameter Error)
:
Indicates that the measurement could not be performed due to internal conflicting
parameter settings.
A good approach to localize the conflicting settings is to start with a reset or preset
or even restore the factory default settings. Then reconfigure the measurement
step by step and check when the error occurs for the first time.
If you need assistance to localize the conflicting parameter settings, please contact
Rohde & Schwarz (see
http://www.service.rohde-schwarz.com
3.5.2 General Measurement Settings
The commands valid for all WCDMA measurements are divided into the groups listed
below.
3.5.2.1
Signal Routing
The following commands configure the scenario, select the input path for the measured
signal and define an external attenuation value.
ROUTe:WCDMa:MEAS<i>:SCENario:SALone
..................................................................620
ROUTe:WCDMa:MEAS<i>:SCENario:CSPath
.................................................................. 620
ROUTe:WCDMa:MEAS<i>:SCENario:MAPRotocol
........................................................... 621