WCDMA Multi Evaluation Measurement
R&S
®
CMW-KG4xx/-KM4xx/-KS4xx
622
User Manual 1173.9657.02 ─ 11
Firmware/Software:
V2.0.10
Manual operation:
See
CONFigure:WCDMa:MEAS<i>:RFSettings:EATTenuation
<RFInputExtAtt>
Defines an external attenuation (or gain, if the value is negative), to be applied to the
RF input connector.
This command is only relevant for the standalone scenario. For the combined signal
path scenario, use
.
Parameters:
<RFInputExtAtt>
Range:
-50 dB to 90 dB
*RST:
0 dB
Default unit: dB
Example:
See
Specifying General Measurement Settings
Firmware/Software:
V1.0.0.4
Manual operation:
See
"External Attenuation (Input)"
3.5.2.2
Analyzer Settings
The following commands configure the RF input path.
CONFigure:WCDMa:MEAS<i>:RFSettings:ENPower
.........................................................622
CONFigure:WCDMa:MEAS<i>:RFSettings:UMARgin
........................................................ 623
CONFigure:WCDMa:MEAS<i>:CARRier<c>:BAND
........................................................... 623
CONFigure:WCDMa:MEAS<i>:RFSettings:CARRier<c>:FREQuency
..................................624
CONFigure:WCDMa:MEAS<i>:RFSettings:DCARrier:SEParation
....................................... 625
CONFigure:WCDMa:MEAS<i>:RFSettings:ENPower
<ExpNomPower>
Sets the expected nominal power of the measured RF signal.
This command is only relevant for the standalone scenario. For the combined signal
path scenario, use:
●
●
Command Reference
135-1095-0799