Remote Control Commands
R&S
®
NRX
218
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>:LIMit<DirectionalChannel>:UPPer:STATe
<state>
Enables or disables the monitoring function for the upper limit.
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or
Parameters:
<state>
*RST:
OFF
Manual operation:
See
See
"Forward Upper Limit State, Reflection Upper Limit State"
13.5.1.5
Result Formats and Screenshots
............................................................................................218
......................................................................................... 219
FORMat:SREGister
<sregister>
Specifies the format that is used for the return value of
.
Parameters:
<sregister>
ASCii | BINary | HEXadecimal | OCTal
*RST:
ASCii
FORMat[:READings][:DATA]
[<data,length>, <arg1>]
Specifies how the controller expects numeric data from the R&S NRX.
Parameters:
<data,length>
<REAL,32 | 64>
Floating point numbers as standardized in IEEE 754, 32-bit or
64-bit. If you omit the length, the R&S NRX uses the last used
length.
Example for
REAL,32
format:
#14....<binary float value>....
Example for
REAL,64
format:
#18....<binary float value>....
Measurement Settings and Results