Sensor Configuration
R&S
®
NRX
119
User Manual 1178.5566.02 ─ 07
Frequency dependent offset active ← Primary Sensor Offsets, Secondary Sensor
Offsets dialogs ← Frequency Dependent Offset
Enables or disables the selected table.
If enabled, the measurement results are corrected using the specified offset. If the
exact frequency value is not available in the table, the values of the table are interpola-
ted. If the selected frequency is outside the specified frequency range, the first or last
offset value of the table is used.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:
TABLe[:STATe]
Frequency dependent offset table ← Primary Sensor Offsets, Secondary Sensor
Offsets dialogs ← Frequency Dependent Offset
Selects one of the available offset tables.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:
TABLe:INDex
Edit table name ← Primary Sensor Offsets, Secondary Sensor Offsets dialogs ←
Frequency Dependent Offset
Enter the table name.
Remote command:
Edit table "<table name>" ← Primary Sensor Offsets, Secondary Sensor Offsets
dialogs ← Frequency Dependent Offset
Opens a dialog to edit the selected table.
Remote command:
MEMory:TABLe:FREQuency:POINts?
Export file name, Import file name ← Primary Sensor Offsets, Secondary Sensor
Offsets dialogs ← Frequency Dependent Offset
Edits the filename for export/import.
Export table to file, Import table from file ← Primary Sensor Offsets, Secondary
Sensor Offsets dialogs ← Frequency Dependent Offset
Exports or imports the table specified under
Export file name, Import file name
Gamma Correction
Enables or disables the gamma correction. See also
Correction Settings