Remote Control Commands
R&S
®
NRX
327
User Manual 1178.5566.02 ─ 07
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<state>
ON | OFF
*RST:
OFF
Manual operation:
See
See
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:TABLe:
INDex
<value>
Selects one of the available offset tables.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Setting parameters:
<value>
Range:
1 to 10
*RST:
1
Usage:
Setting only
Manual operation:
See
"Frequency dependent offset table"
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:TABLe[:
STATe]
<state>
Enables or disables the selected table.
If enabled, the measurement results are corrected using the specified offset. If the
exact frequency value is not available in the table, the values of the table are interpola-
ted. If the selected frequency is outside the specified frequency range, the first or last
offset value of the table is used.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Setting parameters:
<state>
*RST:
OFF
Configuring Sensors