![R&S 1424.6721.02 Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/rands/1424-6721-02/1424-6721-02_user-manual_774880063.webp)
Remote Control Commands
R&S
®
NRP18S-xx
63
User Manual 1178.3686.02 ─ 05
TRIGger:SLOPe
<slope>
Effective only if
INTernal
or
EXTernal
.
Determines which edge of the envelope power, with internal triggering, or increasing
voltage, with external triggering, is used for triggering.
Parameters:
<slope>
POSitive | NEGative
POSitive
Rising edge
NEGative
Falling edge
*RST:
POSitive
TRIGger:SOURce
<source>
Selects the source for the trigger event detector.
Parameters:
<source>
HOLD | IMMediate | INTernal | BUS | EXTernal | EXT1 |
EXTernal1 | EXT2 | EXTernal2
See
Chapter 8.5.2.3, "Trigger Sources"
*RST:
IMMediate
TRIGger:SYNC:PORT
<sync_port>
Selects the external connection for the sync output of the sensor. For more information,
see
.
Parameters:
<sync_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
TRIGger:SYNC:STATe
<state>
Usually used in combination with
ON
.
If enabled, blocks the external trigger bus as long as the sensor remains in the mea-
surement state. Thus, ensures that a new measurement is only started after all sen-
sors have completed their measurements.
Make sure that the number of repetitions is the same for all sensors involved in the
measurement. Otherwise, the trigger bus is blocked by any sensor that has completed
its measurements before the others and has returned to the idle state.
Parameters:
<state>
*RST:
OFF
Controlling the Measurement